DEPUY IRELAND RECLAIM MON REV STD STEM 17 HO; HIP FEMORAL STEM Back to Search Results
Catalog Number 101854517
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/19/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that this patient's right hip was revised due to dislocation.Patient's primary reclaim monobloc stem had rotated within femur, so it was replace by a longer reclaim monobloc stem.Patient's primary femoral head and acetabular liner were also replaced by dual mobility components to further stabilize the patient's hip.Two weeks prior to this surgery, the patient underwent a surgical procedure to convert failed (cut out) trochanteric nail to a total hip using depuy synthes components.The trochanteric nail had been used a few weeks prior to repair a trochanteric femoral fracture the patient had sustained.All three surgical procedures where performed by the doctor at the hospital.Doi: (b)(6) 2026.Dor:(b)(6) 2026.Affected side: right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECLAIM MON REV STD STEM 17 HO
Common Device Name
HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY IRELAND
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key24225135
Report Number1818910-2026-02034
Device Sequence Number13107733
Product Code LZO
UDI-Device Identifier10603295533801
UDI-Public(01)10603295533801(17)340831(10)4586940
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K221462
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number101854517
Device Lot Number4586940
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/19/2026
Supplement Date Received by Manufacturer02/09/2026
Initial Report FDA Received Date01/30/2026
Supplement Report FDA Received Date02/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/07/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTRX +4 10D 40IDX56OD; DLT TS CER HD 12/14 40MM +5.0
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexMale
-
-