Model Number DVFB1D4
Medical Device Problem Codes Premature Discharge of Battery (1057); Failure to Interrogate (1332); Reset Problem (3019)
Health Effect - Clinical Codes Cardiac Arrest (1762); Pneumonia (2011); Ventricular Fibrillation (2130); Cardiogenic Shock (2262); Syncope/Fainting (4411)
Date of Event 01/05/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the patient presented to the hospital experiencing cardiac arrest, pneumonia, lactic acidosis and cardiogenic shock.In addition, the patient was in ventricular fibrillation (vf) for approximately 40 minutes and was unresponsive.Cardiopulmonary resuscitation (cpr) was administered.The implantable cardioverter defibrillator (icd) had potentially reached recommended replacement time (rrt) status unexpectedly.There was suspected premature battery depletion, and two electrical resets occurred.Telemetry was no longer able to be established.A lead impedance warning triggered due to undefined high rv defibrillator coil impedance.The device and lead remain in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was further reported that the device was explanted and replaced and the lead was capped and replaced.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.The returned device indicated shorting/low impedance in the battery.A nalysis of the hybrid revealed the device was fully functional and operated under normal current drain when powered with an external supply.The device passed all manual therapy delivery testing.There was no evidence of a high current drain condition, and no hybrid anomalies were observed.Analysis of the battery revealed plated lithium was found to initiate from the anode, breaching the anode separator bag, and contacting the cathode tab.Based on this, the premature battery depletion was determined to be the result of a plated lithium short between the anode (negative polarity) and the cathode tab (positive polarity).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIA AF MRI XT VR SURESCAN
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key24225430
Report Number3004209178-2026-01577
Device Sequence Number9366055
Product Code LWS
UDI-Device Identifier00643169566422
UDI-Public(01)00643169566422(17)180528
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P980016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/28/2018
Device Model NumberDVFB1D4
Device Catalogue NumberDVFB1D4
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2026
Supplement Date Received by Manufacturer02/03/2026
05/21/2026
Initial Report FDA Received Date01/30/2026
Supplement Report FDA Received Date02/18/2026
06/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/07/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1305-2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6935M62 LEAD
Outcome Attributed to Adverse Event Life Threatening; Hospitalization; Required Intervention;
Patient Age43 YR
Patient SexMale
Patient Weight80 KG
-
-