| Model Number |
DVFB1D4 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Failure to Interrogate (1332); Reset Problem (3019)
|
| Health Effect - Clinical Codes |
Cardiac Arrest (1762); Pneumonia (2011); Ventricular Fibrillation (2130); Cardiogenic Shock (2262); Syncope/Fainting (4411)
|
| Date of Event |
01/05/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented to the hospital experiencing cardiac arrest, pneumonia, lactic acidosis and cardiogenic shock.In addition, the patient was in ventricular fibrillation (vf) for approximately 40 minutes and was unresponsive.Cardiopulmonary resuscitation (cpr) was administered.The implantable cardioverter defibrillator (icd) had potentially reached recommended replacement time (rrt) status unexpectedly.There was suspected premature battery depletion, and two electrical resets occurred.Telemetry was no longer able to be established.A lead impedance warning triggered due to undefined high rv defibrillator coil impedance.The device and lead remain in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was further reported that the device was explanted and replaced and the lead was capped and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was returned and analyzed.The returned device indicated shorting/low impedance in the battery.A nalysis of the hybrid revealed the device was fully functional and operated under normal current drain when powered with an external supply.The device passed all manual therapy delivery testing.There was no evidence of a high current drain condition, and no hybrid anomalies were observed.Analysis of the battery revealed plated lithium was found to initiate from the anode, breaching the anode separator bag, and contacting the cathode tab.Based on this, the premature battery depletion was determined to be the result of a plated lithium short between the anode (negative polarity) and the cathode tab (positive polarity).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|