• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. ETHLN BLK 18IN 7-0 S/A P-1 PRM; SUTURE, NONABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. ETHLN BLK 18IN 7-0 S/A P-1 PRM; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number 1696G
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: - when did the alleged deficiency occur (removal from package / during handling prior to use on patient/ during passage through tissue / during tying / post-op)? if different, please explain.--during passage through tissue - please clarify if all 12 devices displayed in the mentioned attachment demonstrated the alleged deficiency.--yes, all 13 devices displayed in the mentioned attachment demonstrated the alleged deficiency.They have been reported to ethicon, in addition to this complaint report, the corresponding reference numbers are (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.The problem of the suture pulling off the needle happened in surgery.Total 5 products occurred needle pull off issue.Changed to another one to complete the surgery.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 3/3/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803,.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found.H6 component code: c22 - photo analysis.Additional information: d9, h3, h6.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to eth for evaluation.Visual inspection revealed that one empty labeled winding former, and one detached needle were received for analysis.Product code 1696g.During the visual inspection of the returned needle, the swage and attachment area was noted to be as expected.The barrel hole was examined under magnification, and no suture remnant was observed.In addition, the suture was not returned for evaluation; therefore, the cause of the reported event could not be determined.As part of eth quality process all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what caused the reported complaint.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Photo analysis: this is an analysis for a photo submitted to ethicon for evaluation.During the visual analysis, the following was observed: the photo shows twelve overwrap packets.Within each packet a winding former was present along with a detached needle and suture that pertain to product code 1696.Based on the photo review, the event reported is confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHLN BLK 18IN 7-0 S/A P-1 PRM
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key24228617
Report Number2210968-2026-01175
Device Sequence Number15698360
Product Code GAR
UDI-Device Identifier10705031002159
UDI-Public(01)10705031002159(17)280930(10)TLMAHU
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1696G
Device Lot NumberTLMAHU
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2026
Supplement Date Received by Manufacturer02/06/2026
Initial Report FDA Received Date01/30/2026
Supplement Report FDA Received Date03/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-