| Catalog Number |
ROB10024 |
| Medical Device Problem Code |
Application Program Problem: Parameter Calculation Error (1449)
|
| Health Effect - Clinical Code |
Laceration(s) (1946)
|
| Date of Event |
01/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during a cori assisted tka surgery, the (1) real intelligence cori showed that the 5/2 cut block was 3 4 mm off of plan after putting in.Surgeon have been using cori console for a while, and the block was not that far off.Also, when the surgeon went to burr all, the cori console took additional distal femur bone.Checkpoints were double checked and they were not the issue.It took over 2 mm extra of distal and the femur did not fit well.Surgeon manually downsized and recut because it was so messed up.It is unknown if there was any surgical delay.The procedure was resumed with a change in surgical technique (manual procedure).No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Search Alerts/Recalls
|