Catalog Number ESR920
Medical Device Problem Codes Fluid/Blood Leak (1250); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/15/2026
Type of Reportable Event Serious Injury
Event or Problem Description
According to the available information the device was removed and replaced due to broken tubing malfunction.
 
Additional Manufacturer Narrative
As no lot number was provided, a review of the device history record, complaint history database, non-conformances and capas could not be completed.The device was not returned for evaluation.Without the benefit of analyzing the device, quality cannot comment on the condition of the device.Should additional information prompt us to alter or supplement any information or conclusions contained in this report, a follow-up report will be submitted.
 
Event or Problem Description
According to the additional available information, the implant was removed and replaced due to broken tubing above the pump causing a leak.
 
Additional Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.Corrections: b3 and g3.
 
Event or Problem Description
According to the additional available information, the implant was removed and replaced due to broken tubing above the pump causing a leak.
 
Additional Manufacturer Narrative
Titan otr pump was received for evaluation.Abrasion was noted on both exhaust tubes and the inlet tube of the pump.A separation, surrounded by abrasion, was noted on the longer exhaust tubing of the pump.This is a site of leakage.Based on examination of the returned product, it was concluded that while in-vivo both the exhaust tubes and inlet tube of the pump had overlapped and abraded against one another.This positioning, in combination with device usage over time, could contribute to sufficient stress to separate the longer exhaust tubing of the pump.A separation of this type could then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN INFLATABLE PENILE PROSTHESIS FAMILY
Common Device Name
PENILE PROSTHESIS, INFLATABLE
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer (Section G)
MINNEAPOLIS
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key24233592
Report Number2125050-2026-00167
Device Sequence Number15696010
Product Code FHW
UDI-Device Identifier05708932487689
UDI-Public5708932487689
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P000006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/31/2016
Device Catalogue NumberESR920
Device Lot Number3524786_ESR9202400
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date02/02/2026
Supplement Report FDA Received Date02/03/2026
04/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/14/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR