• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. BIPAP S/T C-SERIES; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. BIPAP S/T C-SERIES; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Model Number DS1060TS
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/29/2026
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.A bipap s/t c series device was returned to a third-party service center with no initial alleged complaint.During the evaluation of the device at the third-party service center, the device was visually inspected, and evidence of foam degradation/particles was found inside the blower kit.Additionally, the service technician also noted dust contamination, and the evaluation of the device data identified unrelated error codes.These error codes are consistent with normal device operation and expected use conditions and are not considered reportable under medical device reporting requirements.No evidence was identified to suggest that these error codes contributed to or caused the reported event.The device was scrapped by the service center after the evaluation was completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP S/T C-SERIES
Common Device Name
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key24235054
Report Number2518422-2026-003744
Device Sequence Number10188566
Product Code MNS
UDI-Device Identifier00606959012561
UDI-Public(01)00606959012561
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K102465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Distributor
Initial Reporter Occupation 505
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 02/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberDS1060TS
Device Catalogue NumberDS1060TS
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/29/2026
Initial Report FDA Received Date02/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/23/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1973-2021
Patient Sequence Number1
Patient SexUnknown
-
-