| Catalog Number |
367977 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® sst¿, an unspecified number of tubes are bowed preventing the automated instrument from aspirating the sample properly.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
There were multiple 510k numbers reported to be involved.The information is as follows: g.4.Pma / 510(k)#: k230855.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd had not received any samples or photos for investigation.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: bowed.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® sst¿, an unspecified number of tubes are bowed preventing the automated instrument from aspirating the sample properly.There was no health impact or consequences reported.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® sst¿, an unspecified number of tubes are bowed preventing the automated instrument from aspirating the sample properly.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental mdr is submitted to document that bd will no longer consider the specific malfunction identified below to be a reportable mdr malfunction for the specified product family.Consistent with fda's "medical device reporting for manufacturers" guidance (november 8, 2016, section 2.15), bd determined that this issue has not caused or contributed to a death or serious injury in the previous two years and that the likelihood of a death or serious injury occurring as a result is remote.Bd previously notified fda of this determination on january 8, 2026.Product family: bd vacutainer® sst¿ blood collection tubes
specific malfunction: tubes / extenders with molding defects (non-cosmetic) that can be used
this mdr is being submitted as a representative notification for the product family and specific malfunction identified above and applies to all applicable catalog numbers within that product family associated with this issue.This determination applies only to the issue identified above when it occurs independently and does not otherwise meet mdr reporting requirements.Bd will continue to submit mdrs for reportable deaths, serious injuries, and reportable malfunctions, including events involving multiple event types where mdr reporting criteria are met.Future occurrences of this issue will no longer be submitted as standalone malfunction mdrs.Bd will continue to monitor complaint and adverse event data related to this issue and will reassess mdr reportability should new information indicate that reporting criteria may be met.
|
| |
|