| Catalog Number |
106697 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that an external fluid leak was observed from the drain bag of a prismaflex m100 set approximately three (3) hours after the start of continuous renal replacement therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: the actual device was not available; however, a video of the sample was provided for evaluation.The video was reviewed and showed a leak at the level of the drain bag sealing near the tube.The reported condition was verified.The cause of the condition was determined to be a supplier manufacturing issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|