MAUDE Adverse Event Report: VANTIVE US HEALTHCARE LLC PRISMAFLEX SET; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VANTIVE US HEALTHCARE LLC PRISMAFLEX SET; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 106697
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/07/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an external fluid leak was observed from the drain bag of a prismaflex m100 set approximately three (3) hours after the start of continuous renal replacement therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
H11: the actual device was not available; however, a video of the sample was provided for evaluation.The video was reviewed and showed a leak at the level of the drain bag sealing near the tube.The reported condition was verified.The cause of the condition was determined to be a supplier manufacturing issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX SET
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
VANTIVE US HEALTHCARE LLC
deerfield IL
Manufacturer (Section G)
VANTIVE HEALTHCARE - MEYZIEU
7, av. lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
510 lake cook rd
deerfield, il, IL 60015
8153072966
MDR Report Key24236320
Report Number8010182-2026-00108
Device Sequence Number13308878
Product Code KDI
UDI-Device Identifier07332414064556
UDI-Public07332414064556
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K041005
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 02/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number106697
Device Lot Number25A0073CB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/07/2026
Supplement Date Received by Manufacturer02/19/2026
Initial Report FDA Received Date02/02/2026
Supplement Report FDA Received Date02/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
8010182-12/17/25-001-C
Patient Sequence Number1
Patient SexUnknown
-
-