BECKMAN COULTER, INC AU5800 CLINICAL CHEMISTRY; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
|
Back to Search Results |
|
| Model Number |
AU5800-10 CHEMISTRY ANALYZER AU5800 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported obtaining high sodium patient results on their au5800 cell 1 clinical chemistry analyzer.There was no report of change to treatment, injury, or death associated with this event.The customer did not provide patient data or demographics.
|
| |
|
Additional Manufacturer Narrative
|
|
The beckman field service engineer (fse) evaluated the au5800 clinical chemistry analyzer and determined the defective buffer valves, which were replaced to resolve the issue.The fse performed calibration and quality control with acceptable results.The fse verified the analyzer returned to normal operation.Section a2, a3, a4, and a5: information not provided by customer.Beckman coulter internal identifier is (b)(4).
|
| |
|