Model Number FX819T
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code Hydrocephalus (3272)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: no product received to date.Should relevant additional information/investigation results become available, an additional medwatch report will be submitted.
 
Event or Problem Description
It was reported that a m.Blue plus (#fx819t) was implanted.According to the complainant, the shunt system caused a blockage.The patient underwent a revision procedure.No patient complications were reported as a result of the revision procedure.The complained product was not yet sent to the manufacturer for investigation.
 
Additional Manufacturer Narrative
Visual inspection: during the investigation, no significant deformations or damages of the valves were determined.Permeability test: a permeability test has shown that both valves are permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical position.The results show that the valves are operating within the specified tolerances in their respective relevant position.Adjustment test: the valves were tested and both valves are adjustable in all pressure settings.Braking force and brake function test: the brake functionality test has shown that the brake function operates as expected and the braking force is within the given tolerance.Internal inspection: after dismantling of the valves, no organic deposits were found.Results: first of all, we would like to point out that the condition of the product upon delivery was not ideal for the investigation.The plastic container holding the liquid and the products had already developed a slight mold growth, and the components may have already been affected by it.This influences the results of the investigation, as the integrity of the products may have been compromised.Based on the investigation results, we can determine visible deposits in all components of the shuntsystem.The deposits in the catheter caused the temporary blockage of the shuntsystem.All other components operated within specification at the time of the investigation.Deposits caused by substances naturally present in the patient's body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic material can affect the integrity of the valve.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
The complained product was sent to the manufacturer and the investigation has been completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.BLUE PLUS SYS W/PED.CONTROL RESERVOIR
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key24240433
Report Number3004721439-2026-00040
Device Sequence Number10175170
Product Code JXG
UDI-Device Identifier04041906504823
UDI-Public(01)04041906504823(11)241029(17)291027(10)20078449
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K192266
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX819T
Device Catalogue NumberFX819T
Device Lot Number20078449
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/23/2026
Supplement Date Received by Manufacturer01/23/2026
Initial Report FDA Received Date02/02/2026
Supplement Report FDA Received Date03/26/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-