| Model Number |
H938173 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that eighteen (18) non-vented high-volume inlets had particulate matter in the inlet.This was observed before use.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, h3, h4, h6(update codes), h11.H4: the lot was manufactured in february 2025.H11: eighteen (18) unopened samples were received for evaluation.Visual inspection was performed on all the samples and particulate matter was found either on the outside of the tubing or the spike or the white connector or inside the sealed packaging.The reported condition was verified.The cause of the issue could not be determined; however, the cause was possibly inherent to contamination during the manufacturing or packaging process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|