• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-032
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported on (b)(6) 2026, a 32mm amplatzer septal occluder was chosen for implant using 12f amplatzer trevisio delivery system.During the procedure, while implanting, it was observed that the occluder did not conform to its desired shape and took form of a cobra deformation.The deformed device was never released from the delivery cable.There was no interaction with the cardiac structures during deployment, and no angulation/kink were noticed in the delivery system.The procedure was aborted.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.No additional information was provided.
 
Additional Manufacturer Narrative
An event of device deformity was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.Additionally, the lot was reviewed for similar complaints, and there is no indication of a lot-specific product issue.Based on the information received, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key24240461
Report Number2135147-2026-00580
Device Sequence Number9220343
Product Code MLV
UDI-Device Identifier00811806010236
UDI-Public(01)00811806010236(17)260131(10)7830914
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Model Number9-ASD-032
Device Catalogue Number9-ASD-032
Device Lot Number7830914
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2026
Supplement Date Received by Manufacturer03/02/2026
Initial Report FDA Received Date02/02/2026
Supplement Report FDA Received Date03/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-