|
Aremstar auto a-flex device was returned to the third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.During the evaluation of the device at the service center, the device was visually inspected and found evidence of foam degradation.During the evaluation, foam particles were observed, and dust contamination was found in the device.Additionally, the evaluation of the device data identified unrelated error code.This error code is consistent with normal device operation and expected use conditions and are not considered reportable under medical device reporting requirements.No evidence was identified to suggest that these error codes contributed to or caused the reported event.The device was scrapped.
|