Model Number 102956
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/15/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Section a.Patient information there was no patient involvement.Section d4: expiration date and manufacture date (h4) cannot be determined.The primary udi number in d4 is reflective of all available information.No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
 
Event or Problem Description
It was reported that persistent s3 alarm was noted on power on prior to initiating support.The system was switched to back-up console and motor.Given equipment age, a complete system replacement was performed.There was no patient involved.All equipment was being replaced due to age being past max product life.No log files would be provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation conclusion: the centrimag motor was evaluated following the reported event of the system alarming for an s3 alarm.The centrimag motor (serial number: (b)(6)) was not returned for analysis, no log files were provided, and no additional documentation was submitted for review.Provided information indicated that the alarm activated when the system was powered on, prior to initiating support, that no patient was involved with the reported event, and that the console and motor were exchanged.The reported event could not be correlated to an issue with the centrimag motor.The device history record was reviewed and revealed no deviations with manufacturing and qa specifications.The 2nd generation centrimag system operating manual (rev.D) section 4 ¿ "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual (rev.D) section 10 ¿ "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support." the 2nd generation centrimag system operating manual (rev.D) section 12.1 ¿ "appendix i ¿ primary console alarms and alerts" cover all alarms (auditory and visual), including system alarms, and the appropriate actions to take if the issue does not resolve.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG MOTOR, US
Common Device Name
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
fredric gold
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key24241429
Report Number3003306248-2026-00027
Device Sequence Number9220887
Product Code KFM
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K020271
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number102956
Device Catalogue Number102956
Device Lot Number7455555
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/15/2026
Supplement Date Received by Manufacturer02/03/2026
Initial Report FDA Received Date02/02/2026
Supplement Report FDA Received Date02/05/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/16/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-