| Model Number |
SIGLU60A |
| Medical Device Problem Codes |
Misfire (2532); Failure to Form Staple (2579); Failure to Fire (2610); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Tissue Breakdown (2681)
|
| Date of Event |
01/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: sigc60mt; tri 2.0 sigc60mt 60 med thk cartridge; (lot number: n3e0065y) siglu60a; tri 2.0 siglu60a 60 load unit; (lot number: n3h2068y) sigc60mt; tri 2.0 sigc60mt 60 med thk cartridge; (lot number: n3e0065y) sigc60mt; tri 2.0 sigc60mt 60 med thk cartridge; (lot number: n3e0065y) sigphandle; sig power sigphandle handle; (serial number: unknown) sigadaptxl; sig power sigadaptxl linear xl adapter; (serial number: unknown).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a surgical procedure using the loading units and reloads with a powered stapler and xl adapter, the reloads were not able to form proper b staples, staples that opened and could not close fully, staples fell into the abdominal cavity, reloads were not able to fire fully, incomplete staple lines, and the device not being able to fire even with new blades and reloads which resulted to extended surgical time of two hours due to multiple device issues.It was confirmed that the staples were retrieved by hand instruments.Troubleshooting involved replacing the device with new blades and reloads, but the same issues persisted, leading to the use of staplers from another company.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h2, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the instrument did not fire, component disengaged, partial firing and staples were not formed properly.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|