This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Pump received with corroded battery tube, stripped battery cap coin slot(p), cracked belt clip slot, stained end cap sticker, stained address/serial number label and cracked reservoir tube lip.No broken off battery cap contact noted during visual inspection.The original battery cap had a stripped battery at the coin slot.The pump passed the displacement test and the test infusion set and reservoir does lock into place.Cosmetic damage was confirmed battery cap damaged and battery compartment damaged.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|