It was reported that the doctor placed an implant at tooth site 24, it did not achieve primary stability and was removed.Doctor placed a new implant at tooth site 24, and when the patient took rx control doctor saw a sinus membrane perforation at tooth location 24 and decided remove the implant.Patient suffered inflammation.No other implant was placed in the same surgery.This report belong to the implant of the sinus perforation event.
|
|
Zimvie complaint number (b)(4).Zimvie received one (1) tsx37b16, (tsx implant, 3.7mmd, 16mml) for evaluation.Visual evaluation was performed; the implant has signs of use with debris on its internal and external threads.The implant has markings from removal.There was no damage identified or signs of malfunction that would have contributed to the event.Measurements match drawing.Dhr review was completed for the subject lot number 2120023624.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2120023624 for similar events and no other complaint was identified.Review completed utilizing keywords: ¿dental: medical: perforated sinus¿ based on the investigation and risk management file review, the most likely root cause determined from the investigation was customer error and patient factors.Therefore, based on the available information, a device malfunction did not occur.No damage to the implant was identified.The reported event was non-verifiable with all the available information provided.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|