COVIDIEN MANSFIELD TRUCLEAR; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number 7209509
Medical Device Problem Codes Failure to Cut (2587); Mechanical Jam (2983); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/13/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Correction: h6 (removed rfr: adverse event) additional information: g3 additional review of the event and/or additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned.Therefore, this event did not and does not meet the reporting requirements in 21 cfr 803.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during a hysteroscopy polypectomy the soft tissue shaver was not dissecting properly, appeared slow, and seemed blunt.The dense tissue shaver plus was then used, but the situation did not improve.A reusable hand instrument was also tried, resulting in only a very slight improvement.The polyp being removed was over 3 cm in size.Due to the performance of the devices, the procedure had to be completed using bipolar resection, which was not initially available in the hospital.This switch extended the surgical time by one hour.
 
Additional Manufacturer Narrative
D10 concomitant products: 72203012, dense tis shaver plus 72203012 truclear (lot#:5775673), 7209808, 7209808 truclear control unit (serial#: (b)(6)), 7209807, 7209807 truclear handpiece (serial#: (b)(6)), 72205000, control unit 72205000 hysterolux (serial#: (b)(6)), 72205051, hysteroscope seal 72205051 truclear (lot#:9767402) 72205015, procedure kit 72205015 hysterolux (lot#:w5g0126).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR
Common Device Name
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
COVIDIEN MANSFIELD
15 hampshire street
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN MANSFIELD
15 hampshire street
mansfield MA 02048
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24244684
Report Number1282497-2026-00007
Device Sequence Number18201153
Product Code HIH
UDI-Device Identifier10884521744103
UDI-Public(01)10884521744103(17)280324(10)5832568(20)02
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K132015
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7209509
Device Catalogue Number7209509
Device Lot Number5832568
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2026
Supplement Date Received by Manufacturer01/27/2026
Initial Report FDA Received Date02/03/2026
Supplement Report FDA Received Date02/06/2026
Date Device Manufactured03/24/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age36 YR
Patient SexFemale
-
-