| Model Number |
SCD7A39 |
| Medical Device Problem Codes |
Failure to Deliver Energy (1211); Failure to Fire (2610); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during laparoscopic cholecystectomy, the device stopped working and had a red light.It also had no coagulation issue.The battery was switched then the active blade broke off into the patient which had to be retrieved.There was no patient harm/injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the device had a fractured waveguide.The tip of the waveguide was detached and returned.Functional testing found the a ssembled device returned a green light and produced a welcome tone.The assembled device immediately alarmed when activated.The waveguide was inspected under magnification and the origin of the crack was identified.Investigation personnel concluded that the titanium waveguide became cracked from continued use after it may have come in contact with a metallic object.Use after damage can cause the cracked waveguide to break off.It was reported that the device stopped working and had a red light.It also had no coagulation issue and the device component was disengaged.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: contact between the active blade and other metal objects (hemostats, clips, staples, retractors, etc.) may result in product damage, such as a broken blade.Pieces of a broken blade may fall into the surgical cavity causing unintended tissue damage.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during laparoscopic cholecystectomy, the device stopped working and had a red light.It also had no coagulation issue.The battery was switched then in the middle of the procedure the active blade broke off into the patient which had to be retrieved.All the pieces were retrieved without the use of an x-ray or additional medical procedure.The device was not working intermittently prior to the disengagement of the piece.A new device was used to complete the procedure.There was no patient harm/injury.
|
| |
|