| Model Number |
LF1837 |
| Medical Device Problem Codes |
Energy Output Problem (1431); Intermittent Energy Output (4025)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a sleeve gastrectomy the ligasure blunt tip sealer divider did not coagulate short vessels despite a complete fusion cycle, resulting in inadequate or partial sealing with evidence of energy delivery and end tones heard.There was no re-grasp alert.The procedure was extended by 5-10 minutes.The tissue being sealed was a short vessel with an approximate thickness of 7 mm, and the jaws were clean as it was the beginning of the procedure.Between 5-10 good seals occurred before the issue.Another device was used.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the device has inadequate or partial sealing and it stopped working.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|