| Model Number |
406082 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
This lead was explanted due to dislodgement.Should additional information become available, this file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.The lead is currently not available for analysis.No conclusion can be drawn based on available information at the moment.The file is closed.The investigation will be re-opened should additional data become available.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.Upon receipt, the lead under complaint was found capped 32.5 cm proximal to the lead tip.The distal fragment was received for analysis.An extraction aid was found on the distal fragment.The proximal fragment was not returned.Additionally, the insulation was found damaged 22.5 cm proximal to the lead tip.The damage probably occurred during the explantation procedure.Further analysis of the returned lead fragment did not reveal any irregularity that might have contributed to the reported dislodgement.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
|
| |
|