| Model Number |
IMPLANT SYSTEM, SUTURELOC |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 13-jan-2026, a sales representative reported via phone that during a meniscal root repair procedure on (b)(6) 2026, an ar-4551 sutureloc meniscal root repair kit experienced two issues.First, the knee scorpion needle broke while passing the suture; the fragment was retrieved.Second, a portion of the passport button cannula broke during placement in the patient; this fragment was also retrieved.The case was delayed approximately 15 minutes, requiring additional anesthesia.No patient harm was reported.Additional information was received on 19-jan-2026: the ar-12990 knee scorpion device was used with the knee scorpion needle during the procedure.No damage was reported to the ar-12990 knee scorpion device.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to applying excessive mechanical forces upon insertion.
|
| |
|