• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH URETEROSCOPE 7.3/10.4 FR. X 430 MM, 0°, ANGLED OCULAR, 6.4 FR. CHANNEL; RIGID URETEROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH URETEROSCOPE 7.3/10.4 FR. X 430 MM, 0°, ANGLED OCULAR, 6.4 FR. CHANNEL; RIGID URETEROSCOPE Back to Search Results
Model Number WA2UR32A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was observed that during the device evaluation, the rigid ureteroscope had a fracture at the connecting section of the scope main body where the stopcock was installed.There was no patient involvement.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: component failure of the scope main body.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETEROSCOPE 7.3/10.4 FR. X 430 MM, 0°, ANGLED OCULAR, 6.4 FR. CHANNEL
Common Device Name
RIGID URETEROSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key24246107
Report Number9610773-2026-00701
Device Sequence Number9618352
Product Code FGB
UDI-Device Identifier04042761086578
UDI-Public04042761086578
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K200369
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberWA2UR32A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/06/2026
Initial Report FDA Received Date02/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/04/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-