• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION THE Q-TRAK SUBCUTANEOUS ELECTRODE; IMPLANTABLE ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION THE Q-TRAK SUBCUTANEOUS ELECTRODE; IMPLANTABLE ELECTRODE Back to Search Results
Model Number 3010
Medical Device Problem Codes Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Read Input Signal (1581)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 11/29/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) exhibited oversensing, led to inappropriate electric shock.Evaluate the vector for programming was recommended.Currently, this product remains in service and no additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
The product was already updated.This product remains in service and has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.With the available information, boston scientific concluded the clinical observation of oversensing is a known inherent risk of the product and is noted within the instructions for use (ifu), as well as the product's risk documentation.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.This device remains in service.As such, physical analysis has not been conducted in our laboratory.However, a labeling review and risk review were conducted.These reviews determined that the clinical observation of oversensing is a known inherent risk of the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance with labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.This oversensing has been observed with this product previously and is a known inherent risk associated with its use and is documented in the labeling of this product.
 
Event or Problem Description
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) exhibited oversensing, led to inappropriate electric shock.Evaluate the vector for programming was recommended.Currently, this product remains in service and no additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THE Q-TRAK SUBCUTANEOUS ELECTRODE
Common Device Name
IMPLANTABLE ELECTRODE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
229 avenida fabricante
san clemente CA 92672
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key24247334
Report Number2124215-2026-06050
Device Sequence Number9657486
Product Code LWS
UDI-Device Identifier00802526563201
UDI-Public(01)00802526563201(17)161201(10)A117471
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P110042/S000
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/01/2016
Device Model Number3010
Device Catalogue Number3010
Device Lot NumberA117471
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/15/2026
Supplement Date Received by Manufacturer03/25/2026
Initial Report FDA Received Date02/03/2026
Supplement Report FDA Received Date03/30/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
-
-