MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9750TFX20A
Medical Device Problem Code Gradient Increase (1270)
Health Effect - Clinical Codes Chest Pain (1776); Dyspnea (1816)
Date of Event 10/29/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation is underway.
 
Event or Problem Description
As reported by an edwards lifesciences field clinical specialist, approximately 3 years and 6 months post implantaiton of a 20mm sapien 3 ultra valve in the aortic position, the valve was explanted.Patient was symptomatic with dyspnea on exertion and mild substernal chest pressure.Echo showed severe prosthetic aortic valve stenosis with a mean gradient of 45mmhg across the aortic valve.
 
Additional Manufacturer Narrative
A supplemental mdr is being submitted due to engineering evaluation findings.Sections: g3, g6, h2, h6 type of investigation, investigation findings, investigation conclusions have been updated.The event reported is anticipated in the risk management documentation for transcatheter heart valve procedures.A previous investigation into this type of event is captured in an edwards lifesciences technical summary and applies to this complaint.Additional assessment of the failure mode is not required at this time.The device was not returned for evaluation.Due to the unavailability of the complaint device, engineering was unable to perform any visual inspection, functional testing, or dimensional analysis.Per the technical summary, the instructions for use (ifu), training manuals and current risk mitigations have been reviewed.No inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Through extensive complaint investigations, stenosis/increased gradient events for the s3, s3u, s3ur valves post-implantation have not been associated with device malfunctions or manufacturing nonconformances.Rather, the root cause for these events have historically been due to patient factors (atherosclerosis, thrombosis, endocarditis, rheumatic fever, pannus formation) and/or procedural factors (under-deployed valve, valve size selection, patient-prosthesis mismatch).The reported increased gradients was confirmed based on available information to date.It is possible that one or more patient/procedural factors identified in the technical summary may have been present and contributed to the complaint event.However, due to insufficient information, a definitive root cause is unable to be determined.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.Since no edwards product defects or labeling deficiencies were identified, no corrective or preventative action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE SYSTEM
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key24247771
Report Number2015691-2026-10761
Device Sequence Number9593481
Product Code NPT
UDI-Device Identifier00690103201314
UDI-Public(01)00690103201314(17)250306
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P140031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/06/2025
Device Model Number9750TFX20A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2025
Supplement Date Received by Manufacturer02/04/2026
Initial Report FDA Received Date02/03/2026
Supplement Report FDA Received Date02/04/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-