As reported by an edwards lifesciences field clinical specialist, approximately 3 years and 6 months post implantaiton of a 20mm sapien 3 ultra valve in the aortic position, the valve was explanted.Patient was symptomatic with dyspnea on exertion and mild substernal chest pressure.Echo showed severe prosthetic aortic valve stenosis with a mean gradient of 45mmhg across the aortic valve.
|
|
A supplemental mdr is being submitted due to engineering evaluation findings.Sections: g3, g6, h2, h6 type of investigation, investigation findings, investigation conclusions have been updated.The event reported is anticipated in the risk management documentation for transcatheter heart valve procedures.A previous investigation into this type of event is captured in an edwards lifesciences technical summary and applies to this complaint.Additional assessment of the failure mode is not required at this time.The device was not returned for evaluation.Due to the unavailability of the complaint device, engineering was unable to perform any visual inspection, functional testing, or dimensional analysis.Per the technical summary, the instructions for use (ifu), training manuals and current risk mitigations have been reviewed.No inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Through extensive complaint investigations, stenosis/increased gradient events for the s3, s3u, s3ur valves post-implantation have not been associated with device malfunctions or manufacturing nonconformances.Rather, the root cause for these events have historically been due to patient factors (atherosclerosis, thrombosis, endocarditis, rheumatic fever, pannus formation) and/or procedural factors (under-deployed valve, valve size selection, patient-prosthesis mismatch).The reported increased gradients was confirmed based on available information to date.It is possible that one or more patient/procedural factors identified in the technical summary may have been present and contributed to the complaint event.However, due to insufficient information, a definitive root cause is unable to be determined.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.Since no edwards product defects or labeling deficiencies were identified, no corrective or preventative action is required.
|