Jd a healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced an excessive vault, refractive surprise, blurred vision, lens rotation, glare/ haloes, significant reduction of ica, and visual disturbances.The lens was exchanged with a shorter length lens, and the problems were resolved.There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
|
|
H3: device evaluation: the lens was returned dry, stuck on paper with debris found on the lens.Visual inspection found the lens haptic torn and debris on the lens.Dimensional inspections found the lens to be within specifications.Functional inspections found the lens did not meet original values measured at the time of manufacturing.H6: type of investigation 3331: device history record (dhr) review: based on the results of the investigation, all released devices from the associated work order(s), including the suspected device, have been manufactured within established process parameters and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim: (b)(4).
|