|
MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
SMHX400 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
09/11/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: capsular contracture.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with mentor memorygel xtra breast implants experienced capsular contracture baker grade iii/iv on the left side post-operatively, which was confirmed during a physical exam.As a result, the patient underwent device removal and replacement.
|
| |
|
Additional Manufacturer Narrative
|
|
On feb 16, 2026, the mentor failure analysis lab received the device for evaluation.Device evaluation summary: the product was returned to mentor for evaluation.Mentor then conducted visual inspection, leak test and microscopic examination of the returned device.Visual analysis of the returned device determined that the breast implant was found to have bubbles within gel.Leak testing was performed, in accordance with mentor procedures, and two tears (a,b) were found on the posterior aspect of the implant measuring all tears (a,b) approximately 0.1 cm.No additional areas of gel exposure were found.Microscopic examination was performed on the edges of the ruptures (a, b) , and parallel striations were found in the whole area of the tears.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.Based on the information currently available, microscopic examination of the returned product indicates that the implant could have been damaged during or subsequent to implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps or scissors to contact the device during the implantation or other surgical procedures.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|