| Catalog Number |
HAR9F |
| Medical Device Problem Codes |
Material Separation (1562); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a hernia repair procedure the tip fell off from the device.No other information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 2/3/2026.D4 batch #: unknown.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.The expiration date is currently not available.Therefore, the full udi is currently not available.Attempts are being made to obtain the following information.To date, no response has been provided.If further details are received at a later date, a supplemental medwatch will be sent.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot/batch was not provided; therefore, the manufacturing records evaluation could not be performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 3/11/2026.D4 batch #: y70243.Corrected data: d4 udi #.Investigation summary- the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the device was returned with the tissue pad melted, and 100% present.The blade was not tip broke off as reported by customer.No suspect diversion were identified.The device was connected to a test handpiece and a gen11, and the device did activate during functional testing.Probable causes of tissue pad damage are applying pressure between the instrument blade and tissue pad without having tissue between them.Avoid activating the blade without tissue between the blade and tissue pad to prevent damage to the tissue pad.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch y70243, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|