|
VARIAN MEDICAL SYSTEMS, INC. INTELLIBLATE XIMITRY PROBE; INTELLIBLATE XIMITRY PROBE ASSEMBLY, 13G X 15CM
|
Back to Search Results |
|
| Model Number |
IB-XPA-1315 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Air Embolism (1697)
|
| Date of Event |
12/30/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During a microwave ablation procedure using an intelliblate ib-xpa-1315 probe, the system triggered a cooling system alarm approximately one minute after ablation began at 60 w.Upon responding to the alarm, the physician performed imaging and identified a bolus of air within the liver and in the right ventricle of the heart.The physician immediately stopped the system and removed the probe, at which point it was observed that the distal tip of the probe had detached and remained inside the patient.Subsequent imaging confirmed that the detached tip was retained within the liver and appeared broken just after the feeding point, exhibiting a slight burnt aspect.Because the cooling pump continued to operate following the tip detachment, air entered through the broken section of the probe and migrated into the liver vasculature, ultimately reaching the right ventricle.The patient received antibiotic treatment and was monitored in the icu for two days without further incident.The detached tip remains in the patient.No long-term health consequences are anticipated.Varian has requested additional information; the damaged probe will be returned for analysis.The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer received a report regarding an incident that occurred during a microwave ablation procedure involving an intelliblate ibxpa1315 probe.Approximately one minute after ablation began at 60 w, the system issued a cooling system alarm.In response to the alarm, the physician performed imaging, which identified the presence of air within the liver and in the right ventricle of the heart.The physician immediately stopped the system and withdrew the probe.Upon removal, it was noted that the distal tip of the probe had detached and remained inside the patient.Subsequent imaging confirmed that the detached tip was retained within the liver.The component appeared fractured just distal to the feeding point and exhibited a slight burnt appearance.Based on the report, continued operation of the cooling pump following the tip detachment allowed air to enter the probe through the break and migrate into the vasculature of the liver, ultimately reaching the right ventricle.The patient received antibiotic therapy and was monitored in the icu for two days without further incident.The detached tip remains in the patient per the treating physician¿s assessment.No long-term health consequences are anticipated.Varian has requested additional information from the user facility.The damaged probe will be returned to the manufacturer for analysis.The investigation is ongoing, and a follow-up report will be submitted when the investigation is completed or if additional information is received.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data provided - device evaluation: h6, h11.The manufacturer received a report regarding an incident that occurred during a microwave ablation procedure involving an intelliblate ib xpa 1315 probe.Approximately one minute after ablation began at 60 w, the system issued a cooling system alarm.In response to the alarm, the physician performed imaging, which identified the presence of air within the liver and in the right ventricle of the heart.The physician immediately stopped the system and withdrew the probe.Upon removal, it was noted that the distal tip of the probe had detached and remained inside the patient.Subsequent imaging confirmed that the detached tip was retained within the liver.The returned component appeared fractured just distal to the feeding point and exhibited a slight burnt appearance.Based on the report, continued operation of the cooling pump following the tip detachment allowed air to enter the probe through the break and migrate into the vasculature of the liver, ultimately reaching the right ventricle.The patient received antibiotic therapy and was monitored in the icu for two days without further incident.The detached tip remains in the patient per the treating physician¿s assessment.No long-term health consequences are anticipated.The returned probe was visually inspected.Inspection confirmed that the distal tip was detached and no longer secured to the probe shaft.Based on the observed damage, the detachment is consistent with insufficient coolant flow to the probe during use.The specific cause of the insufficient coolant flow could not be definitively confirmed.However, one possible contributing factor could be improper seating of the coolant fluid spike within the coolant fluid bag, which may allow air to enter the coolant line and reduce coolant flow.Under such conditions, the system is designed to generate error code 5, indicating a potential probe overtemperature condition and prompting the user to verify proper spike insertion and adequate coolant bag temperature.The ximitry probe instructions for use (p1071010003c, p.17) instruct users to insert the coolant fluid spike into the coolant fluid bag using a twisting motion and to verify there are no leaks.Once the system is fully primed, air should not be present in the coolant lines.Improper preparation of the coolant bag may result in insufficient coolant flow, which could lead to probe overheating and subsequent tip detachment.The ximitry probe ifu (p1071010003c, p.30) includes a warning that insufficient coolant fluid flow can result in unintended thermal injury and instructs that, if the pump is active and fluid is not flowing, the tubing may be damaged or blocked and the probe assembly should be replaced.Additionally, the intelliblate console ifu (p1071008003c, p.51) troubleshooting guidance instructs users who observe visible air gaps in the fluid line to verify that the coolant fluid spike is fully inserted into the coolant fluid bag, check for and remove any kinks in the fluid lines, and replace the assembly if air gaps persist.
|
| |
|
Event or Problem Description
|
|
During a microwave ablation procedure using an intelliblate ib-xpa-1315 probe, the system triggered a cooling system alarm approximately one minute after ablation began at 60 w.Upon responding to the alarm, the physician performed imaging and identified a bolus of air within the liver and in the right ventricle of the heart.The physician immediately stopped the system and removed the probe, at which point it was observed that the distal tip of the probe had detached and remained inside the patient.Subsequent imaging confirmed that the detached tip was retained within the liver and appeared broken just after the feeding point.The returned portion of the probe exhibited a slight burnt aspect, consistent with thermal exposure.Because the cooling pump continued to operate following the tip detachment, air entered through the broken section of the probe and migrated into the liver vasculature, ultimately reaching the right ventricle.The patient received antibiotic treatment and was monitored in the icu for two days without further incident.The detached tip remains in the patient.No long-term health consequences are anticipated.
|
| |
|
Search Alerts/Recalls
|
|
|