MENTOR TEXAS MENTOR SMOOTH ROUND HIGH PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3503630 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
01/12/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and an associated nonconformance was found.The nonconformance will be handled through mentor¿s quality system.Reason for device explant and/or reoperation: breast prosthesis deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation revision procedure with a mentor smooth round high profile saline breast prosthesis that deflated post implantation.Deflation of the patient¿s left breast prosthesis was diagnosed via a healthcare professional¿s exam.As a result, the patient underwent bilateral explantation and replacement with mentor memorygel xtra breast implant gel breast prostheses.
|
| |
|
Additional Manufacturer Narrative
|
|
On 26-feb-2026, the mentor failure analysis lab received the device for evaluation.On 05-mar-2026, mentor completed its investigation on the suspect medical device.Device evaluation summary: according to the information received, the patient experienced a deflation in the breast implant.The product was returned to mentor for evaluation.Mentor conducted visual inspection and microscopic examination of the returned device.Visual analysis of the returned device revealed that the breast implant had a crease/fold on the anterior view.In addition, a tear was observed within the crease/fold, measuring approximately 2.8 cm.Microscopic examination was performed and no instrument damage was observed.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures as it can result in a deflation in a later time.Breast implants may also simply wear out over time.The contralateral device was also received (lot number 9578063).The patient did not report any issue with this device.Therefore, no further analysis of the contralateral device is required.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|