| Model Number |
CI-1601-04 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
01/16/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Event or Problem Description
|
|
The recipient reportedly experienced a tip fold-over.A ct scan confirmed the tip fold-over.The recipient reportedly was a poor performer.The recipient's device was explanted.The recipient was re-implanted with another cochlear device.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information section: h.6.Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed and silicone damage was observed on the top and bottom cover of the device.These are believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock was verified.The condition of the electrode prevented an electrical test from being performed.The device passed some of the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information section: h.6.Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed and silicone damage was observed on the top and bottom cover of the device.These are believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock was verified.The condition of the electrode prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|