• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS Back to Search Results
Model Number DXR00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Disturbances (2140)
Date of Event 12/30/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that following implantation of preloaded multifocal intraocular lenses (iols), model dxr00v, the patient experienced shadowed vision and glare intolerance in the right eye.Lens was was subsequently explanted during a secondary surgical procedure.No incision enlargement, vitrectomy, sutures, surgical delays, or medications outside of standard care were required.The patient¿s best corrected visual acuity remained 20/20 both before and after the procedures.The patient reported temporary visual impairment, including an inability to drive at night after the primary surgery.Replacement lens implanted in right eye was non¿johnson & johnson products.Post operative refraction for the replacement lens has not yet been measured.No additional information was provided.
 
Additional Manufacturer Narrative
Section h3: the device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of records including device history and complaint trending will be performed.Upon completion of the review, if there is any further relevant information, a supplemental medwatch will be filed.Section h6: health effect- impact code: 4619 - glare surgical intervention required.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: february 3, 2026.Section h3: device evaluated by manufacturer: yes.Device evaluation: visual inspection revealed that the lens was cut into three pieces and coated in viscoelastic residue.The lens was cleaned and inspected; no further issues were identified.During the product evaluation it was confirmed that the physical characteristics of the lens matched a dxr00v model lens.No issues that could cause or contribute to the complaint issues reported were identified during product evaluation.The ¿lens cut¿ observed during product evaluation could not be confirmed to be related to the manufacturing or design process.Conclusion: based on the complaint investigation results, no product deficiency or product malfunction could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key24254826
Report Number3012236936-2026-000024
Device Sequence Number15540242
Product Code HQL
UDI-Device Identifier05050474761827
UDI-Public(01)05050474761827(17)271216
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberDXR00V
Device Catalogue NumberDXR00VU215
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2026
Supplement Date Received by Manufacturer02/13/2026
Initial Report FDA Received Date02/03/2026
Supplement Report FDA Received Date03/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/16/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-