MASIMO - 15750 ALTON PKWY RADICAL-7; OXIMETER Back to Search Results
Model Number 9500
Medical Device Problem Code Data Problem (3196)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Other text: device evaluation is anticipated.When the device is evaluated or new information is obtained, a follow-up report will be submitted.Health effect impact code: no adverse event, no patient impact.E1.Initial reporter zip code exceeded allowable field length, initial reporter zip code is as follows: (b)(6).
 
Event or Problem Description
The customer reported the radical-7s' "mode changed unintentionally." per additional information from the customer, "the mode automatically switches from pediatric mode to neonatal mode." there was no patient impact or consequence reported.
 
Event or Problem Description
The customer reported the radical-7s' "mode changed unintentionally." per additional information from the customer, "the mode automatically switches from pediatric mode to neonatal mode." there was no patient impact or consequence reported.
 
Additional Manufacturer Narrative
Other text: the returned device was evaluated.During evaluation the device passed all visual and functional testing.Throughout testing and after each power cycle, the device remained in pediatric mode.The device was confirmed to be functioning as designed.Health effect impact code: no adverse event; no patient impact.(b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADICAL-7
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
rishi jain
15750 alton pkwy
irvine, CA 92618
MDR Report Key24254905
Report Number3019388613-2026-00033
Device Sequence Number8327357
Product Code DQA
UDI-Device Identifier00843997008358
UDI-Public(01)00843997008358(11)221029
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K212161
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 01/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9500
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2026
Supplement Date Received by Manufacturer03/04/2026
Initial Report FDA Received Date02/03/2026
Supplement Report FDA Received Date03/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-