W. L. GORE & ASSOCIATES, INC. GORE® DRYSEAL FLEX INTRODUCER SHEATH; INTRODUCER, CATHETER
|
Back to Search Results |
|
| Catalog Number |
GDSF2233 |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Insufficient Device Problem Information (3190); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Ischemia (1942); Paresis (1998); Movement Disorder (4412); Reduced Blood Flow (4598)
|
| Date of Event |
01/12/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2025, the patient underwent endovascular procedure using a gore® excluder® thoracoabdominal branch endoprosthesis (tambe) device system (unknown).On (b)(6) 2025, the patient experienced an acute blood loos anemia and on (b)(6) 2025 had a treatment using a unit of packed red blood cells (prbc).On (b)(6) 2025, it was noticed that the patient had a lower extremity reduced perfusion.It was reported that it was potentially related to a gore® dryseal flex introducer sheath.The patient was treated by 5000 units heparin iv, a left common femoral artery endarterectomy and 1 cm bovine patch.On (b)(6) 2025, there was an acute kidney injury noticed and was treated by coil x 2 placement in perforated branch and intentional occlusion of the left renal artery with a vascular plug and stent.The issue was related procedures.On (b)(6) 2025, the patient had spinal cord ischemia and a bilateral lower extremity motor loss was noted.The patient was treated by the lumbar drain.According to the site, it was a procedure related event.On (b)(6) 2025, there were left renal infarct and multiple acute right renal cortical infarcts.Procedure related.No treatment was required.On (b)(6) 2025, the patient had a shock- spinal cord ischemia and cardiogenic.Procedure related.No treatment was required.On (b)(6) 2025, the patient had a multi system organ failure and expired.
|
| |
|
Additional Manufacturer Narrative
|
|
C20: a review of the manufacturing records for the device could not be conducted because the serial/lot number remains unknown.The device discarded at the facility and is not available for analysis.It was reported that the patient had a multi system organ failure and expired.Please note that a gore® excluder® thoracoabdominal branch endoprosthesis (tambe) device system was covered under the (b)(4) and was reported separately.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, the patient underwent endovascular procedure using a gore® excluder® thoracoabdominal branch endoprosthesis (tambe) device system.A gore® dryseal flex introducer sheath was used for access.On (b)(6) 2025, the patient experienced an acute blood loos anemia and on (b)(6) 2025, had a treatment using a unit of packed red blood cells (prbc).On (b)(6) 2025, it was noticed that the patient had lower extremity reduced perfusion.It was reported that it was potentially related to a gore® dryseal flex introducer sheath.The patient was treated by 5000 units heparin iv, a left common femoral artery endarterectomy and 1 cm bovine patch.Based on information provided, the patient lost pulse in the left lower extremity (lle) after 22fr sheath was pulled.This was noted by loss of doppler tone at the ankle and no motor response on the mep monitoring.Upon exposure of the left femoral artery and deep femoral artery, it was noted that there was an old healed thrombosed pseudoaneurysm in the lateral circumflex femoral artery (lcfa).Embolic debris was removed.Subject was heparinized throughout the procedure.On (b)(6) 2025, there was an acute kidney injury noticed and was treated by coil x 2 placement in perforated branch and intentional occlusion of the left renal artery with a vascular plug and stent.The issue was related procedures.On (b)(6) 2025, the patient had spinal cord ischemia and a bilateral lower extremity motor loss was noted.The patient was treated by the lumbar drain.According to the site, it was a procedure related event.On (b)(6) 2025, there were left renal infarct and multiple acute right renal cortical infarcts.Procedure related.No treatment was required.On (b)(6) 2025, the patient had a shock- spinal cord ischemia and cardiogenic.Procedure related.No treatment was required.On (b)(6) 2025, the patient had a multi system organ failure and expired.On (b)(6) 2026, the physician mentioned that the patient¿s death and many of the aes are not related to the devices themselves but related to the procedure or past medical history.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: description summary was updated.B7: medical history.D: device information was added.H6.Code c19: a review of manufacturing records of the device indicated the lot met pre-release specifications.The device remains implanted and is not available for analysis.According to the physician, the patient¿s death and many of the adverse events occurred were not related to the gore devices themselves but related to the procedure or past medical history.Please note that a gore® excluder® thoracoabdominal branch endoprosthesis (tambe) device system was covered under the case- (b)(4) and was reported separately.
|
| |
|
Search Alerts/Recalls
|
|
|