| Model Number |
CL-936 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 - concomitant products: cl-936, cl-936 polysorb* 1 vio 90cm gs12 x36, (lot# a5c1940y); cl-936, cl-936 polysorb* 1 vio 90cm gs12 x36, (lot# a5c1940y); cl-936, cl-936 polysorb* 1 vio 90cm gs12 x36, (lot# a5c1940y); cl-936, cl-936 polysorb* 1 vio 90cm gs12 x36, (lot# a5c1940y).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during the hemiarthroplasty, a total of five sutures experienced needle detachment while being removed from the retainer.An additional new suture was used without issue.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection of the returned product noted one needle.No suture was returned for evaluation.The needle was returned disengaged from the suture.Upon microscopic inspection of the needle, normal crimp and swage marks were noted.Suture material was observed to be present in the swage, indicating the suture broke off at the swage.A visual inspection of the returned photos noted two foil pouches and a violet suture inside a plastic clear bag was observed in the images provided.The foil packages were noted to be torn in the upper right quadrants.T was reported that there was needle detachment while being removed from the retainer.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|