• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN SURGICAL POLYSORB; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN SURGICAL POLYSORB; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC Back to Search Results
Model Number CL-936
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 - concomitant products: cl-936, cl-936 polysorb* 1 vio 90cm gs12 x36, (lot# a5c1940y); cl-936, cl-936 polysorb* 1 vio 90cm gs12 x36, (lot# a5c1940y); cl-936, cl-936 polysorb* 1 vio 90cm gs12 x36, (lot# a5c1940y); cl-936, cl-936 polysorb* 1 vio 90cm gs12 x36, (lot# a5c1940y); medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during the hemiarthroplasty, a total of five sutures experienced needle detachment while being removed from the retainer.An additional new suture was used without issue.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted of the returned product noted one suture and one needle.The needle was returned disengaged from the suture.Upon microscopic inspection, instrument marks were noted near the swaged end of the needle.Suture material was observed to be present in the swage, indicating the suture broke off at the swage.A visual inspection of the returned photos showed two foil pouches and a violet suture contained within a clear plastic bag.The foil packages were noted to be torn in the upper right quadrants.Functional testing on the opened complaint device was precluded as the needle was returned detached.It was reported that there was needle detachment while being removed from the retainer.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur if the needle is grasped near the swaged end or the tip and could cause bends, breaks, or damage the connection between the needle and suture.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: in handling this or any other suture material, care should be taken to avoid damage from handling.Avoid crushing or crimping damage due to application of surgical instruments, such as forceps or needle holders.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYSORB
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC
Manufacturer (Section D)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24255206
Report Number1219930-2026-00682
Device Sequence Number15851332
Product Code GAM
UDI-Device Identifier10884521043527
UDI-Public(01)10884521043527(17)300228(10)A5C1940Y(20)01
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K963253
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCL-936
Device Catalogue NumberCL-936
Device Lot NumberA5C1940Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/14/2026
Supplement Date Received by Manufacturer04/02/2026
Initial Report FDA Received Date02/03/2026
Supplement Report FDA Received Date04/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/12/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient Age83 YR
Patient SexFemale
-
-