| Model Number |
QVM2-N-HM |
| Medical Device Problem Code |
No Visual Prompts/Feedback (4021)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The quantum ventilation module units were functioning as expected; however, when the level sensor was connected to the qvm, it failed to operate properly.When the protected/safe level safety was engaged, the led on the level sensor did not illuminate when fluid was present.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is being returned and an investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The quantum ventilation module units were functioning as expected; however, when the level sensor was connected to the qvm, it failed to operate properly.When the protected/safe level safety was engaged, the led on the level sensor did not illuminate when fluid was present.
|
| |
|
Additional Manufacturer Narrative
|
|
Preliminary results showed that with no level sensor connected, the system displays level 2 on qws and spt.When the level sensor is connected, the system displays level 0, and no light is emitted from the sensor.This indicates a potential fault with the pcba - fpl.Root cause and investigation summary with follow in final report.
|
| |
|
Event or Problem Description
|
|
The quantum ventilation module units were functioning as expected; however, when the level sensor was connected to the qvm, it failed to operate properly.When the protected/safe level safety was engaged, the led on the level sensor did not illuminate when fluid was present.
|
| |
|
Additional Manufacturer Narrative
|
|
Preliminary results showed that with no level sensor connected, the system displays level 2 on qws and spt.When the level sensor is connected, the system displays level 0, and no light is emitted from the sensor.This indicates a potential fault with the pcba - fpl.The device has now been returned and the root cause and investigation summary will follow in a final report.
|
| |
|