| Model Number |
40-405-1 |
| Medical Device Problem Code |
Output above Specifications (1432)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/22/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from a healthcare provider regarding a generator.It was reported that during preventive maintenance (pm) on the generator, bipolar output failure occurred, with the generator unable to produce wattage within tolerance and output wattage measured at more than double the maximum tolerance during testing.Impedance measurement was 100 ohms, wattage measurement was 110 watts, and output measurement was 280 watts, with other similar tests confirming the output issue.There was no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3,h6): no parts have been received by the manufacturer for evaluation.Codes b17, c20 <(>&<)> d15 are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|