| Catalog Number |
72205109 |
| Medical Device Problem Codes |
Fracture (1260); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a surgery, it was noticed that the tip of the incisor blade stopped oscillating.On inspection, the blade had fractured at the internal joint where the blade moves (no pieces need to be retrieved from inside the patient as all of them were contained inside the device).The procedure was resumed, without any delay, using a similar sn back up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10 h3, h6: the reported device was received for evaluation.A visual evaluation showed the device was returned with original packaging with the batch number of the complaint on the label.The color scheme of the device matches the print.There are no visible defects until the inner blade is pulled out of the outer blade.The inner blade has fractured at the proximal end of the curve in the tube shaft.The distal end of the inner blade remains inside the outer blade.Bio debris is present.No functional testing can be conducted since there is damage hindering the inspection/evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences (force/side-loading the device during use or contact with another source).Based on this investigation, the need for corrective action is not indicated.
|
| |
|
Search Alerts/Recalls
|
|
|