| Model Number |
2110 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One device was received for evaluation.Visual inspection was performed.Functional testing noted a faulty downstream occlusion sensor.The sensor was replaced.The device then passed all testing criteria.The service history review had no indication that the complaint was related to a service of the device within the review period.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was alarming "cassette not attached properly.Reattach cassette".There was no reported patient involvement.The device evaluation noted a faulty downstream occlusion sensor.
|
| |
|