MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SYNCHRONY; Cochlear Implant Back to Search Results
Model Number Mi1200 SYNCHRONY
Medical Device Problem Code Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
The user showed affected channels since (b)(6) 2025 and also noticed that the hearing performance is affected.Clinical follow-up is ongoing with the clinical team.
 
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
The user showed affected channels since (b)(6) 2025 and also noticed that the hearing performance is affected.Clinical follow-up is ongoing with the clinical team.
 
Additional Manufacturer Narrative
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is suspected.However, to determine an exact root cause a device investigation of the explanted device is necessary.Further technical checks are planned but no date has been scheduled yet.
 
Event or Problem Description
The user showed affected channels since (b)(6) 2025 and also noticed that the hearing performance is affected.The user was reimplanted.
 
Additional Manufacturer Narrative
Conclusion: damage to the active electrode, likely caused by minute device mobility, was determined to have led to device failure over time.Other mechanical damages found are attributable to the removal surgery.The problems given in the recipient report appear to match the damage found.This is a final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRONY
Common Device Name
Cochlear Implant
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
+43()577885626
MDR Report Key24258328
Report Number9710014-2026-00112
Device Sequence Number15427237
Product Code MCM
UDI-Device Identifier09008737093952
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/17/2025
Device Model NumberMi1200 SYNCHRONY
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/08/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2026
Supplement Date Received by Manufacturer02/02/2026
02/02/2026
Initial Report FDA Received Date02/04/2026
Supplement Report FDA Received Date03/13/2026
09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age7 YR
Patient SexFemale
-
-