• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US DURALOC CURVED CUP IMPACTOR; HIP INSTRUMENTS : IMPACTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US DURALOC CURVED CUP IMPACTOR; HIP INSTRUMENTS : IMPACTORS Back to Search Results
Catalog Number 224408000
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/21/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint#: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the device is cross threaded.It did not happen in surgery.There were no direct effect from a case/ surgeon.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> according to the information provided: "it was reported that the event(cross threaded) did not happen in surgery".The product was returned to depuy synthes for evaluation.Visual inspection revealed the threaded cup connector was not returned for evaluation.Device body had signs of usage on its overall surface and a crack was found around the inner metal pin slot.No other anomaly was identified on the evidence provided.The handle may crack or break due to the impact forces generated when the surgeon strikes the handle's anvil with a heavy object while positioning the shell.These forces are transmitted from the anvil directly into the handle.Furthermore of the repeated use, the observed condition of the device can be consistent with thermal expansion of the sterilization process.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.Missing component condition may occur during the cleaning/sterilization process, particularly when the components are assembled or disassembled.As these procedures are often repeated, there is an increased risk of components going missing.However, without concrete evidence or further information, it is not possible to determine a root cause.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was not confirmed as the observed condition of the duraloc curved cup impactor would have not contributed to the complained device issue.¿ based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot
=
> a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code a1007 provided is not a valid finished goods lot number.Corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURALOC CURVED CUP IMPACTOR
Common Device Name
HIP INSTRUMENTS : IMPACTORS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer Contact
elba bello
700 orthpaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key24258751
Report Number1818910-2026-02409
Device Sequence Number7975733
Product Code HWA
UDI-Device Identifier10603295107477
UDI-Public(01)10603295107477(10)A1007
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number224408000
Device Lot NumberA1007
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/21/2026
Supplement Date Received by Manufacturer02/10/2026
02/27/2026
Initial Report FDA Received Date02/04/2026
Supplement Report FDA Received Date02/12/2026
03/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2007
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-