ALLERGAN (COSTA RICA) Inspira Textured Silicone Gel Filled Breast Implant; prosthesis, breast, noninflatable, internal, silicone gel-filled
|
Back to Search Results |
|
| Catalog Number |
N-TRF295 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Patient, through other company representative, reported "rupture".This record is for the left side.The device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Additional, changed, and/or corrected data: a2, b3, b5, b6, d1, d2a, d2b, d4, d6b, g4, h4, h6.
|
| |
|
Event or Problem Description
|
|
Patient, through other company representative, reported "rupture".Healthcare professional later, through distributor, reported "rupture".This record is for the left side.The device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation:
per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
- rupture: observed broken device through microscopic inspection assessed as unidentified (tear) opening
and missing piece of shell assessed as inconclusive.No further actions are required as it is not related to the manufacturing process.No additional observations performed on the device.No further actions are required.Additional, changed, and/or corrected data: a5, b5, d4, d9, g2, h3, h6.
|
| |
|