It was reported that, during a tha surgery, upon implantation of the r3 cup, one (1) r3 offset impactor tip broke off from the impactor or handle.All pieces were recovered and the wound was irrigated with copious amounts of irrigation.The procedure was resumed, without any delay, with a change in surgical technique.No further complications were reported.
|
|
Internal complaint reference: (b)(4).H11: the associated device was returned and evaluated.A visual inspection of the returned device finds heavy wear from use with the device returned fractured into pieces, the edges scraped and gouged and the surfaces damaged.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history of the previous 12 months revealed similar events for the listed part number, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|