• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. BIO-COMP, SWIVELOCK C, FT, 4.75X19.1MM; BIO SOFT TISSUE FIXATN FASTNR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BIO-COMP, SWIVELOCK C, FT, 4.75X19.1MM; BIO SOFT TISSUE FIXATN FASTNR Back to Search Results
Model Number BIO-COMP, SWIVELOCK C, FT, 4.75X19.1MM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 15th jan 2026, it was reported by an arthrex's employee via an email that for 1 unit of ar-2324bcct, suture anchor, biocomposite swivelock® c vented, 4.75 x 19.1 mm, with fibertape® loop (blue), its anchor broke during insertion.This was detected during a reconstruction of the anterior cruciate ligament of the knee joint on (b)(6) 2025.The procedure was completed successfully by using another new ar-2324bcct.There were no patient harm and no secondary procedure needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-COMP, SWIVELOCK C, FT, 4.75X19.1MM
Common Device Name
BIO SOFT TISSUE FIXATN FASTNR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key24259979
Report Number1220246-2026-00417
Device Sequence Number9851047
Product Code MAI
UDI-Device Identifier00888867026841
UDI-Public(01)00888867026841(11)241202(17)281130(10)15355206
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K101823
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 05/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBIO-COMP, SWIVELOCK C, FT, 4.75X19.1MM
Device Catalogue NumberAR-2324BCCT
Device Lot Number15355206
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/15/2026
Supplement Date Received by Manufacturer05/19/2026
Initial Report FDA Received Date02/04/2026
Supplement Report FDA Received Date05/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-