MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG CONNECTED OR HUB BASE SYSTEM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG CONNECTED OR HUB BASE SYSTEM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 0240200200
Medical Device Problem Codes No Display/Image (1183); Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that all the monitors shut off during the case, and they lost video for quite some time.Not sure to the extent that it remained off, but the case was aborted due to malfunction.No harm as a result of the camera box, but case was aborted and rescheduled.There is also suspicion to the hub possibly shutting off and restarting mid-case as to an explanation for why all monitors lost signal.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: no live video output.Probable root cause: cables, connectors, sources, or sinks, capture card, motherboard, power supply, sdc firmware, over-heating (air duct, fans, heat sinks, dust, vents), sdc application software, clarity package, clarity algorithm, manufacturing defects (process, workmanship), use error, cybersecurity.The reported failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that all the monitors shut off during the case, and they lost video for quite some time.Not sure to the extent that it remained off, but the case was aborted due to malfunction.No harm as a result of the camera box, but case was aborted and rescheduled.There is also suspicion to the hub possibly shutting off and restarting mid-case as to an explanation for why all monitors lost signal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG CONNECTED OR HUB BASE SYSTEM
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
arsalan rizvi
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key24260873
Report Number0002936485-2026-00057
Device Sequence Number7975788
Product Code GCJ
UDI-Device Identifier07613327601695
UDI-Public07613327601695
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K241401
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0240200200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/07/2026
Supplement Date Received by Manufacturer03/23/2026
Initial Report FDA Received Date02/04/2026
Supplement Report FDA Received Date04/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;