| Model Number |
M635WU60400 |
| Medical Device Problem Code |
Failure to Seal (4070)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a device leak occurred.A left atrial appendage (laa) closure procedure was performed, and a 40mm watchman flx pro closure device was implanted on (b)(6) 2025.During the procedure at the time of release, the physician noted a small leak.No further intervention was performed at that time and the patient was discharged.The physician planned for closure of the leak, which was successfully performed on (b)(6) 2026 with a non-boston scientific product.There were no patient complications, and the patient is expected to fully recover.
|
| |
|
Event or Problem Description
|
|
It was reported that a device leak occurred.A left atrial appendage (laa) closure procedure was performed, and a 40mm watchman flx pro closure device was implanted on (b)(6) 2025.During the procedure at the time of release, the physician noted a small leak.No further intervention was performed at that time and the patient was discharged.The physician planned for closure of the leak, which was successfully performed on (b)(6) 2026 with a non-boston scientific product.There were no patient complications, and the patient is expected to fully recover.
|
| |
|
Additional Manufacturer Narrative
|
|
With all the available information, boston scientific (bsc) concludes: device technical analysis the watchman flx? pro remains implanted; therefore, returned device analysis could not be completed.Device history record review a review was completed of the device history records (dhr) for the watchman flx? pro, part # m635wu60400, batch # 0036958460.The device was processed through all normal operational conditions and passed all acceptance activities.The dhr review did not identify any deviations or non-conformances that could have contributed to the event.Labeling review there was evidence to suggest that the device was used in a manner inconsistent with the labeling.Watchman flx? pro instructions for use (ifu) lists potential complications, risks and adverse events that may be associated with the use of this device which include implant did not seal risk review a risk review was completed and confirmed that the event of implant did not seal was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
|
| |
|
Search Alerts/Recalls
|