| Catalog Number |
368884 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® lh (lithium heparin) 68 i.U.Plus blood collection tubes, an unspecified number of tubes have issues filling.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd did not receive any samples or photos for investigation.Therefore, functional tests were conducted using 20 retained samples and all samples were found to be within specification.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: underfill.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® lh (lithium heparin) 68 i.U.Plus blood collection tubes, an unspecified number of tubes have issues filling.There was no health impact or consequences reported.
|
| |
|