• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TANDEM DIABETES CARE TANDEM MOBI INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY; ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TANDEM DIABETES CARE TANDEM MOBI INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY; ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP Back to Search Results
Model Number 1004000
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that intermittent occlusion alarms occurred.Reportedly, the customer had been using the same cartridge and infusion set for over 3 days on the mobi pump.Tandem customer technical support informed customer that the cartridge and infusion is indicated for use with the mobi pump for 3 days.Customer changed cartridge and infusion set to address the issue.There was no reported adverse impact to the customer¿s blood glucose level.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TANDEM MOBI INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
Common Device Name
ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP
Manufacturer (Section D)
TANDEM DIABETES CARE
12400 high bluff drive
san diego CA 92130
Manufacturer Contact
john connell
12400 high bluff drive
san diego, CA 92130
858 366-6900
MDR Report Key24262028
Report Number3013756811-2026-22562
Device Sequence Number8010236
Product Code QFG
UDI-Device Identifier00389152271910
UDI-Public(01)00389152271910(22)1014027(11)240516
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K233044
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1004000
Device Catalogue Number1012719
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2026
Supplement Date Received by Manufacturer04/13/2026
Initial Report FDA Received Date02/04/2026
Supplement Report FDA Received Date05/05/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/16/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HUMALOG - INSULIN; HUMALOG - INSULIN; TANDEM - CARTRIDGE; UNOMEDICAL - INFUSION SET; UNOMEDICAL - INFUSION SET
Patient Age46 YR
Patient SexFemale
-
-