• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYMATESE EVOLYSSE SMOOTH; IMPLANT, DERMAL, FOR AESTHETIC USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYMATESE EVOLYSSE SMOOTH; IMPLANT, DERMAL, FOR AESTHETIC USE Back to Search Results
Catalog Number EVO-SMTH-01
Medical Device Problem Codes Off-Label Use (1494); Improper or Incorrect Procedure or Method (2017); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Obstruction/Occlusion (2422); Swelling/ Edema (4577)
Date of Event 11/28/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The filler was injected into the patient and is not available for return.Based on the reported information, this event is consistent with a vascular occlusion which is a known serious adverse event documented in the labeling.This can occur due to unintended injection into a blood vessel or injection of a large amount of filler near a blood vessel causing compression of the blood vessel.Both occlusion and compression may result in reduced or cessation of blood supply to tissues.Intervention is required to prevent serious and permanent complications.Additionally, the product was used off-label in the lip region.We cannot rule out that this may have contributed to the reported event.The production records for the lot were reviewed, and no deviations were found during production that would relate to the reported event; the device met all specifications.Eus-20260108-14[2] and 3549.
 
Event or Problem Description
On (b)(6), 2025 the hcp reporting injecting evolysse smooth into both the upper and lower lips of a patient.By november 28, 2025, the patient experienced yellow crusting on her lips.The patient was prescribed antivirals with no improvement.She was then prescribed the first medrol does steroid pack without improvement.The patient was then treated with bactroban (antibiotic) and a second medrol steroids dose pack without improvement.The product was dissolved and resulted in a significant improvement to the patient.However, the patient is experiencing discomfort due to an impaired barrier which is being managed with gabapentin and flexeril.The patient is experiencing referred pain in the jaw and sensitivity in the lips.As of january 05, 2026 the patient's lips have re-epithelialized and symptoms are improving.As of january 27, 2026, the patient has fully recovered and is no longer experiencing any pain symptoms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLYSSE SMOOTH
Common Device Name
IMPLANT, DERMAL, FOR AESTHETIC USE
Manufacturer (Section D)
SYMATESE
zi les troques
chapnost, auvergne-rhone-alpes 69630
FR  69630
Manufacturer (Section G)
SYMATESE
zi les troques
chapnost, auvergne-rhone-alpes 69630
FR   69630
Manufacturer Contact
laurence hermitte
zi les troques
chapnost, auvergne-rhone-alpes 69630
FR   69630
MDR Report Key24263521
Report Number3003672980-2026-00003
Device Sequence Number10669984
Product Code LMH
UDI-Device Identifier03760172161271
UDI-Public(01)03760172161271(17)270331(10)S2251220015
Combination Product (Y/N)Y
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P240022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 02/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberEVO-SMTH-01
Device Lot NumberS2251220015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2026
Initial Report FDA Received Date02/04/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/04/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-